{"created":"2023-05-15T16:43:44.464124+00:00","id":19199,"links":{},"metadata":{"_buckets":{"deposit":"a3812ca5-985c-4e2f-8a6d-079aeada6dfc"},"_deposit":{"created_by":2,"id":"19199","owners":[2],"pid":{"revision_id":0,"type":"depid","value":"19199"},"status":"published"},"_oai":{"id":"oai:nagasaki-u.repo.nii.ac.jp:00019199","sets":["29:68:1348:1354"]},"author_link":["79319","79318","79316","79317","79320"],"item_3_biblio_info_6":{"attribute_name":"書誌情報","attribute_value_mlt":[{"bibliographicIssueDates":{"bibliographicIssueDate":"1993-03-31","bibliographicIssueDateType":"Issued"},"bibliographicPageEnd":"106","bibliographicPageStart":"103","bibliographicVolumeNumber":"6","bibliographic_titles":[{"bibliographic_title":"長崎大学医療技術短期大学部紀要"},{"bibliographic_title":"Bulletin of the School of Allied Medical Sciences, Nagasaki University","bibliographic_titleLang":"en"}]}]},"item_3_description_4":{"attribute_name":"抄録","attribute_value_mlt":[{"subitem_description":"生体インピーダンス法による体脂肪測定器インピーダンスファットメーターSIF-891の測定条件について検討した.その結果,10回連続測定による同時再現性は変動係数が0.81と良好であった.被験者の測定時における姿勢の影響では,上下肢の屈曲および上下肢の開き方による皮膚の接触が測定値に影響を及ぼした.測定部位の皮膚の処置では,アルコール清拭,電極ゲルの処置のいずれを省略しても測定値は低くなった.また,電極装着位置のずれにより電極間の距離が長くなると測定値は高く,短くなると低くなる傾向があった.しかし,立位および座位では測定値に影響はなく,また飲水・摂食・排尿前後の測定では,測定値に一定の傾向はみられなかった.","subitem_description_type":"Abstract"}]},"item_3_description_64":{"attribute_name":"引用","attribute_value_mlt":[{"subitem_description":"長崎大学医療技術短期大学部紀要 = Bulletin of the School of Allied Medical Sciences, Nagasaki University. 1993, 6, p.103-106","subitem_description_type":"Other"}]},"item_3_source_id_10":{"attribute_name":"書誌レコードID","attribute_value_mlt":[{"subitem_source_identifier":"AN10145685","subitem_source_identifier_type":"NCID"}]},"item_3_source_id_7":{"attribute_name":"ISSN","attribute_value_mlt":[{"subitem_source_identifier":"09160841","subitem_source_identifier_type":"ISSN"}]},"item_3_text_62":{"attribute_name":"sortkey","attribute_value_mlt":[{"subitem_text_value":"P00103-P00106"}]},"item_3_version_type_16":{"attribute_name":"著者版フラグ","attribute_value_mlt":[{"subitem_version_resource":"http://purl.org/coar/version/c_970fb48d4fbd8a85","subitem_version_type":"VoR"}]},"item_creator":{"attribute_name":"著者","attribute_type":"creator","attribute_value_mlt":[{"creatorNames":[{"creatorName":"福山, 由美子"}],"nameIdentifiers":[{}]},{"creatorNames":[{"creatorName":"西山, 久美子"}],"nameIdentifiers":[{}]},{"creatorNames":[{"creatorName":"浦田, 秀子"}],"nameIdentifiers":[{}]},{"creatorNames":[{"creatorName":"勝野, 久美子"}],"nameIdentifiers":[{}]},{"creatorNames":[{"creatorName":"大塚, 健作"}],"nameIdentifiers":[{}]}]},"item_files":{"attribute_name":"ファイル情報","attribute_type":"file","attribute_value_mlt":[{"accessrole":"open_date","date":[{"dateType":"Available","dateValue":"2020-12-23"}],"displaytype":"detail","filename":"iryou06_00_16_t.pdf","filesize":[{"value":"397.2 kB"}],"format":"application/pdf","licensetype":"license_note","mimetype":"application/pdf","url":{"label":"iryou06_00_16_t.pdf","url":"https://nagasaki-u.repo.nii.ac.jp/record/19199/files/iryou06_00_16_t.pdf"},"version_id":"0061c722-d0a0-4b86-82ae-b9460e7ce04b"}]},"item_keyword":{"attribute_name":"キーワード","attribute_value_mlt":[{"subitem_subject":"生体インピーダンス","subitem_subject_scheme":"Other"},{"subitem_subject":"体脂肪計","subitem_subject_scheme":"Other"},{"subitem_subject":"測定条件","subitem_subject_scheme":"Other"},{"subitem_subject":"インピーダンスファットメーターSIF-891","subitem_subject_scheme":"Other"}]},"item_language":{"attribute_name":"言語","attribute_value_mlt":[{"subitem_language":"jpn"}]},"item_resource_type":{"attribute_name":"資源タイプ","attribute_value_mlt":[{"resourcetype":"departmental bulletin paper","resourceuri":"http://purl.org/coar/resource_type/c_6501"}]},"item_title":"生体インピーダンス法による体脂肪計の測定条件に関する検討","item_titles":{"attribute_name":"タイトル","attribute_value_mlt":[{"subitem_title":"生体インピーダンス法による体脂肪計の測定条件に関する検討"}]},"item_type_id":"3","owner":"2","path":["1354"],"pubdate":{"attribute_name":"公開日","attribute_value":"2008-07-01"},"publish_date":"2008-07-01","publish_status":"0","recid":"19199","relation_version_is_last":true,"title":["生体インピーダンス法による体脂肪計の測定条件に関する検討"],"weko_creator_id":"2","weko_shared_id":-1},"updated":"2023-05-16T03:27:10.131308+00:00"}